-40%
Julabo MV-4 Heated Water Bath/Circulator - Bad Temperature Sensor
$ 132
- Description
- Size Guide
Description
Julabo MV-4 Heated Water Bath, used, in good cosmetic condition but non-functional. Sold as-is for parts or repair. Firmware Rev 5.0In manual operation mode, the circulator turns on and the unit starts to heat, but then fails with either error E-01 (safety temp exceeded) or E-06 (more than 25° difference between the safety temp sensor and the regular temp sensor).
The keypad and display seem fully functional, and I was able to change and store new temperature settings. I have not tested the remote operation or any other functionality.